BindingDB logo
myBDB logout

Assay Method Information

Assay Name:  HDAC4 Biochemical Assay
Description:  5 μL of each solution of 1:20 diluted compound from above was transferred to a clear bottomed, black, 384-well assay plate using the Bravo or the Janus (384-well MDT head from Perkin Elmer). Using a 16-channel Matrix multi-channel pipette, 35 μL of the working solution of HDAC4 catalytic domain enzyme (0.2 μg/mL in assay buffer) was transferred to the assay plate. The assay was then started by adding 10 μL of 5× (50 μM) substrate to the assay plates using either the Bravo, Janus or 16-channel Matrix multi-channel pipette. The assay plate was then shaken for two minutes on an orbital shaker at 900 rpm (rotations per minute). Next the plate was incubated for 15 minutes at 37° C. The reaction was stopped by adding 25 μL of 3× (30 μM) developer/stop solution to the assay plates using either the Bravo, Janus or a 16-channel Matrix multi-channel pipette. Assay plates were then shaken for 5 minutes on an orbital shaker at 1200 rpm. Next, the assay plates were incubated at 37° C. for 1 hour in a tissue culture incubator. Finally, the fluorescence was measured (Excitation: 355 nm, Emission: 460 nm) using PerkinElmer EnVision in top read mode.
Affinity data for this assay
 

If you find an error in this entry please send us an E-mail
   
    

Home

|

Search

|

Deposit

|

SiteMap

|

About us

|

Email us

|

Info

 
Last update November 1, 2007
©2000 BindingDB. All rights reserved.